Abstract
Objective
This study was undertaken to assess the clinical utility, safety, and tolerability in epilepsy patients of ultra long-term monitoring with a novel subcutaneous electroencephalographic (EEG) device (sqEEG).
Methods
Five patients with drug-resistant focal epilepsy were implanted (one patient bilaterally) with sqEEG. In phase 1, we assessed sqEEG sensitivity for seizure recording by recording seizures simultaneously with scalp EEG in the epilepsy monitoring unit (EMU). sqEEG was scored either visually (v-sqEEG) or by using a semiautomatic algorithm (EpiSight; E-sqEEG). In phase 2, the patients were monitored as outpatients for 3–6 months. sqEEG data were analyzed monthly, evaluating concordance of data obtained by v-sqEEG, E-sqEEG, and patients’ diaries. v-sqEEG data were used to guide treatment adjustments. sqEEG-related side effects were assessed throughout the study.
Results
In phase 1, v-sqEEG detected all seizures recorded in the EMU in all patients, whereas E-sqEEG was as effective in three patients. In the other two patients, E-sqEEG detected only a proportion or none of the seizures, respectively. Sensitivity of E-sqEEG depended on the ictal EEG features. In phase 2, a 100% concordance between E-sqEEG and v-sqEEG in seizure detection was observed for the same three patients as in phase 1. In the other two patients (one implanted bilaterally), effectiveness of E-sqEEG in detecting seizure as compared to v-sqEEG ranged from 0% to 83%. v-sqEEG showed that all patients reported in their diaries fewer seizures than they actually suffered. In four of five patients, v-sqEEG showed that the treatment adjustments had been ineffective or associated with a seizure increment. The only side effect was an infection at the implantation site in one patient.
Significance
The sqEEG system could collect reliable information on seizure activity, thus providing clinically relevant information. Sensitivity of EpiSight in detecting seizures varied across patients, depending on the ictal EEG features. sqEEG ultra long-term monitoring was feasible and well tolerated.
SEP